Last Update : 03/05/2024 - 9:00 Am

Fluke107 Palm-sized Digital Multimeter

rating

Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, re..

  • 1.SYP 1,600,241.SYP

  • Brand: ATC
  • Product Code:18037020
  • Availability:Pre-Order

Storage Code: EC03X0101

Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, relative humidity at 0 % to 75 %. Accuracy specifications take the form of: ±([% of Reading] + [Number of Least Significant Digits])

Accuracy Specifications
AC Volts
(40 Hz to 500Hz)1 Range Resolution Accuracy
6.000 V
60.00 V
600.0 V 0.001 V
0.01 V
0.1 V 1.0 % + 3
DC Volts 6.000 V
60.00 V
600.0 V 0.001 V
0.01 V
0.1 V 0.5 % + 3
AC Millivolts 600.0 mV 0.1 mV 3.0 % + 3
Diode Test2 2.000 V 0.001 V 10 %
Resistance
(Ohms) 400.0 O
4.000 kO
40.00 kO
400.0 kO
4.000 MO
40.00 MO 0.1 O
0.001 kO
0.01 kO
0.1 kO
0.001 MO
0.01 MO 0.5 % + 3
0.5 % + 2
0.5 % + 2
0.5 % + 2
0.5 % + 2
1.5% + 3
Capacitance3 50.00 nF
500.0 nF
5.000 µF
50.00 µF
500.0 µF
1000 µF 0.01 nF
0.1 nF
0.001 µF
0.01 µF
0.1 µF
1 µF 2 % + 5
2 % + 5
5 % + 5
5 % + 5
5 % + 5
5 % + 5
Frequency4
Hz
(10 Hz – 100 kHz) 50.00 Hz
500.0 Hz
5.000 kHz
50.00 kHz
100.0 kHz 0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
0.1 kHz 0.1 % + 3
Duty Cycle4 1 % to 99 % 0.1 % 1 % typical 5
AC Current
(40 Hz to 200 Hz) 4.000 A
10.00 A 0.001 A
0.01 A 1.5 % + 3
DC Current 4.000 A
10.00 A 0.001 A
0.01 A 1.5 % + 3
All AC, Hz, and duty cycle are specified from 1 % to 100 % of range. Inputs below 1 % of range are not specified.
Typically, open circuit test voltage is 2.0 V and short circuit current is <0.6 mA.
Specifications do not include errors due to test lead capacitance and capacitance floor (may be up to 1.5 nF in the 50 nF range).
All AC, Hz, and duty cycle are specified from 1 % to 100 % of range. Inputs below 1 % of range are not specified.
Typical means when the frequency is at 50 Hz or 60 Hz and the duty cycle is between 10 % and 90 %
Function Overload Protection Input Impedance (Nominal) Common Mode Rejection Ratio Normal Mode Rejection Ratio
AC Volts 600 V 1 >10 MO <100 pF 2 >60 dB at dc,
50 Hz or 60 Hz -
AC Millivolts 600 mV >1M, <100 pF >80 dB at 50 Hz or 60 Hz -
DC Volts 600 V 1 >10 MO <100 pF >100 dB at dc,
50 Hz or 60 Hz >60 dB at 50 Hz or 60 Hz
6 x 105 V Hz Max
For mV (AC), input impedance is approximately 1 MO.

Write a review

Note: HTML is not translated!
    Bad           Good